GE MAC 2000 Service Manual page 95

Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

Replacing the Printer Module Assembly
Use the following procedure to replace the printer module assembly.
1.
Insert the new printer module assembly into the bottom cover and snap it into
place.
Ensure the screw holes in the printer module assembly line up properly with
the screw holes in the bottom cover.
2.
Secure the new printer module assembly to the bottom cover with four (4)
screws.
3.
Reattach the chassis.
See step
4.
Reattach the top cover assembly.
See
instructions.
5.
Reinsert the battery.
See step
6.
Continue with the appropriate functional checkout procedure for this FRU.
See
Printer Motor and Gear Assembly
The following procedures describe how to replace the printer motor and gear
assembly.
Removing the Printer Motor and Gear Assembly
Use the following procedure to remove the printer motor and gears from the printer
module assembly.
1.
Before beginning any steps, follow the instructions in
FRU Replacement" on page
2.
Remove the printer module assembly.
See
2053535-003C
1
through step
6
in
"Reattaching the Top Cover Assembly to the Device" on page 81
6
through step
8
in
"Functional Checkout" on page 129
"Removing the Printer Module Assembly" on page 94
MAC™ 2000 ECG Analysis System
"Attaching the Chassis" on page 85
"Replacing the Battery" on page 74
for more information.
73.
Maintenance
for instructions.
for
for instructions.
"Preparing the System for
for instructions.
95