Printer Door
The following procedures describe how to remove and replace the printer door.
Removing the Printer Door
Use the following procedure to remove the printer door.
1.
Before beginning any steps, follow the instructions in
FRU Replacement" on page
2.
Remove the top cover assembly, if it is has not been removed already.
See
3.
Detach the chassis.
See
4.
Remove the printer module assembly.
See
5.
Gently turn the device over so the bottom cover is facing you.
6.
Remove the four screws securing the printer door to the bottom cover, as shown
in the following photograph.
7.
Turn the bottom cover over and remove the dampers on the either side of the
printer door, as shown in the following photograph.