Chapter 4 Configuration; Set Up The Daniel Meterlink; Field Setup Wizard - Emerson Daniel 3410 Series Installation Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Installation manual
DAN-20057315
4
Configuration
After the mechanical and electrical installation is complete use the following to install
MeterLink
and verify meter performance.
4.1

Set up the Daniel MeterLink

Procedure
1. Follow the instructions in the Daniel MeterLink
2. Select File → Program Settings and customize the user-preferences (e.g. User
3. Connect to your meter. If your meter is not shown in the list, select Edit Meter
4. Save the meter configuration file, collect a Maintenance log and Waveforms to
5. Run the Field Setup Wizard.
4.2

Field Setup Wizard

Procedure
1. Use the Field Setup Wizard-Startup in Daniel MeterLink
2. Use General setup to configure the meter's units system (U.S Customary or Metric
3. Set the Frequency/Digital Outputs Sources for either a frequency output or a digital
Model 3418 GUSM
in order to establish connection with the meter to perform final configuration
Meters Quick Start Manual (P/N 3-9000-763) to setup software communications with
the meter.
name, Company name, display units, Liquid Meter volume units and other interface
settings).
Directory and setup the connections properties.
document the meter's "As Found" settings.
that allow proper configuration for your meter (Temperature, Pressure, Meter
Corrections, Meter Outputs, Gas chromatograph setup, Continuous flow analysis
and View local display setup). Selections on this page will affect other configuration
selections.
a) Select Next to continue to General setup.
units) volume units, flow rate time, low flow cutoff, contract hour, enable reverse
flow alarm, set meter time and notepad comments.
a) Select Next to continue to Frequency/Digital Outputs page.
Note
The Meter's Units system configured on the General Page affect the units for the
optional Local Display items.
status.
a) Select the Source for each Frequency/Digital output and select the desired
drive Mode. The Mode options are Open Collector which requires an external
Software for Gas and Liquid Ultrasonic
and select the checkboxes
Configuration
June 2019
83