Procedure
1. Rotate the end caps clockwise fully closing and compressing the end cap seal. Install
2. Install the security seal wire into and through one of the two holes in the end cap.
3. Adjust the security wire, removing all slack and thread into the lead seal.
4. Crimp lead seal and cut wire ends to remove excess wire.
3.6.2
Base enclosure security seals
Use the following instructions to install the security seal wire on the Base Enclosure.
Procedure
1. Install security wire seal into and through the hole in the socket head screw on the
Model 3418 GUSM
the Security latch for each end cap using a 3mm Allen wrench.
a) Choose holes that minimize counterclockwise rotation of the end cap when
the security wire is taut (maximum wire diameter .078 inch; 2.0mm).
Figure 3-22: Transmitter electronics enclosure security seals
A. Transmitter electronics enclosure endcap
B. Security wire seals
Base Enclosure cover (maximum wire diameter .078 inch; 2.0mm).
Electrical installation
June 2019
77