1
2
3
10
9
4.2 Cable Glands and Connectors
1
Figure 10 OPT100 DGA Monitor Cable Glands and Connectors
1
Oil Out: connection for oil return line
2
Oil In: connection for oil intake
3
Ground terminal
4
RS-485: cable gland for RS-485 connection
5
Relay control out: cable gland for relay control to power supply unit
6
DC in: 24 VDC connection from power supply unit
7
Ethernet: external RJ45 connector for permanent Ethernet connection
4
5
8
7
2
3
Figure 9 Inside OPTPSU1 Power Supply Unit
Enclosure
1
Circuit breaker for AC power (F1)
2
Surge arresters
3
Power switch (S1)
4
Power supply
6
5
Relays (3 pcs). Each relay has a LED
that is lit when the relay is active.
6
DC OK LED. If flashing, check DC
wiring.
7
DC out terminal block (X5)
8
Relay control terminal block (X4)
9
Relay output terminal block (X3)
10
Mains power in terminal block (X1)
4
5
Chapter 4 – Electrical Installation
6
7
29