Vaisala Optimus OPT100 User Manual

Vaisala Optimus OPT100 User Manual

Dga monitor
Hide thumbs Also See for Optimus OPT100:
Table of Contents

Advertisement

Quick Links

M211858EN-K
User Guide
Vaisala Optimus ™ DGA Monitor for
power transformers
OPT100

Advertisement

Table of Contents
loading

Summary of Contents for Vaisala Optimus OPT100

  • Page 1 M211858EN-K User Guide Vaisala Optimus ™ DGA Monitor for power transformers OPT100...
  • Page 2 This product contains software developed disclosed to a third party without prior by Vaisala or third parties. Use of the written permission of the copyright holder. software is governed by license terms and Translated documents and translated...
  • Page 3: Table Of Contents

    Table of contents Table of contents About this document..................7 Version information..................7 Related manuals....................8 Documentation conventions................8 Trademarks......................9 Patent notice..................... 9 Product overview..................10 Overview of OPT100 DGA Monitor............... 10 Basic features and options................10 Measured parameters..................10 2.3.1 Total gas pressure measurement............12 2.3.2 Calculating relative moisture saturation inside transformer....12 Notes for normal operation................13...
  • Page 4 OPT100 User Guide M211858EN-K 3.11 Configuring user interface security............... 51 3.12 Configuring Modbus protocol............... 53 3.13 Configuring DNP3 protocol................54 3.14 Configuring IEC 61850 protocol..............56 3.15 Using protocol test mode................57 3.16 Configuring relays..................58 Maintenance....................60 Maintenance safety..................60 Cleaning DGA monitor.................. 60 Maintenance.....................
  • Page 5 Table of contents Technical support................... 101 Warranty......................101 Recycling......................101...
  • Page 6 OPT100 User Guide M211858EN-K List of figures Figure OPT100 front parts with Ground Mounting Set........14 Figure 2 OPT100 rear parts with Ground Mounting Set........15 Figure 3 OPT100 front parts with Wall Mounting Set..........16 Figure 4 OPT100 rear parts with Wall Mounting Set..........17 Figure 5 Cable glands and connectors on OPT100 cabinet........18 Figure 6...
  • Page 7 List of tables List of tables Table Document versions (English)................7 Table 2 Related manuals....................8 Table 3 Applicable patents or applications..............9 Table 4 Measured parameters..................10 Table 5 Calculated parameters..................11 Table 6 Meaning of status LEDs..................21 Table 7 Modbus TCP communication settings............22 Table 8 Default Modbus RTU communication settings.........
  • Page 8 OPT100 User Guide M211858EN-K...
  • Page 9: About This Document

    Chapter 1 – About this document 1. About this document 1.1 Version information Table 1 Document versions (English) Document code Date Description M211858EN-K January 2021 Applicable from software version 1.13.0 onwards. Updated sections: • Total gas pressure measurement (page 12) • Gas level alerts (page 22) •...
  • Page 10: Related Manuals

    OPT100 User Guide M211858EN-K Document code Date Description M211858EN-H July 2020 Applicable from software version 1.11.0 onwards. Added sections: • Protocol test mode (page 24) • Using protocol test mode (page 57) • Configuring autocalibration (page 47) Updated sections: • Autocalibration (page 25) •...
  • Page 11: Trademarks

    Lists tools needed to perform the task. Indicates that you need to take some notes during the task. 1.4 Trademarks Vaisalaâ is a registered trademark of Vaisala Oyj. OPTIMUS ™ is a trademark of Vaisala Oyj. Modbusâ is a registered trademark of Schneider Automation Inc.
  • Page 12: Product Overview

    OPT100 User Guide M211858EN-K 2. Product overview 2.1 Overview of OPT100 DGA Monitor OPT100 is a device for monitoring the internal condition of an electrical transformer using dissolved gas analysis (DGA), also known as fault gas analysis. OPT100 is intended to be installed next to the transformer, and connected to it by two oil lines.
  • Page 13: Table 5 Calculated Parameters

    Chapter 2 – Product overview Parameter Output unit(s) Description Ethane (C Measured from extracted gas Ethylene (C Measured from extracted gas Acetylene (C Measured from extracted gas Carbon monoxide (CO) Measured from extracted gas Carbon dioxide (CO Measured from extracted gas Hydrogen (H Measured in-oil Moisture (H...
  • Page 14: Total Gas Pressure Measurement

    Temperature at location where new relative moisture saturation is needed at -1662.70 A is a water solubility coefficient. Vaisala uses this standard value in all of its oil moisture transmitters. You can also use this equation with an oil-specific coefficient if it has been determined for your oil.
  • Page 15: Notes For Normal Operation

    Chapter 2 – Product overview 2.4 Notes for normal operation In normal operation the DGA monitor performs measurement cycles automatically without any operator involvement. Integration with a SCADA system is recommended for convenient monitoring of measurement results and system status. Status LEDs on the cabinet door indicate system status on a general level, and detailed status is available from the browser- based user interface.
  • Page 16: Parts With Ground Mounting Set

    OPT100 User Guide M211858EN-K 2.5 Parts with Ground Mounting Set Figure 1 OPT100 front parts with Ground Mounting Set Status LEDs OPT100 DGA Monitor Door lock (3 pcs) Power supply unit Door lock Mounting stand Wedge anchors (6 pcs)
  • Page 17: Figure 2 Opt100 Rear Parts With Ground Mounting Set

    Chapter 2 – Product overview Figure 2 OPT100 rear parts with Ground Mounting Set Weather shield Cable glands and oil connections for DGA Monitor Cable glands for power supply unit More information ‣ Dimensions with Ground Mounting Set (page 83)
  • Page 18: Parts With Wall Mounting Set

    OPT100 User Guide M211858EN-K 2.6 Parts with Wall Mounting Set Figure 3 OPT100 front parts with Wall Mounting Set Status LEDs OPT100 DGA Monitor Door lock (3 pcs) Power supply unit Door lock...
  • Page 19: Figure 4 Opt100 Rear Parts With Wall Mounting Set

    Chapter 2 – Product overview Figure 4 OPT100 rear parts with Wall Mounting Set Installation beam (upper) Weather shield Installation beam (lower) Cable glands for DGA Monitor Cradle for power supply unit Cable glands for power supply unit More information ‣ Dimensions with Wall Mounting Set (page 84)
  • Page 20: Opt100 Cable Glands And Connectors

    Relay control to OPTPSU1. Use Vaisala 7 … 12 mm 24 mm cable CBL210539. (0.28 … 0.47 in) DC in DC input from OPTPSU1. Use Vaisala cable 9 … 16 mm 30 mm CBL210544. (0.35 … 0.63 in) Ethernet Ethernet connection. Use shielded outdoor Ethernet cable.
  • Page 21: Optpsu1 Cable Glands And Connectors

    Relay control from DGA monitor. Use 7 … 12 mm 24 mm Vaisala cable CBL210539. (0.28 … 0.47 in) DC out DC output to DGA monitor, 24 V dc, 20 A. 9 … 16 mm 30 mm Use Vaisala cable CBL210544. (0.35 … 0.63 in)
  • Page 22: Interior Parts

    OPT100 User Guide M211858EN-K 2.9 Interior parts Figure 7 Inside OPT100 DGA Monitor cabinet Valve 5 (bleed valve). Must be manually accessed during initialization and uninstallation. Valve 4 Oil block Valve 3 Valve 2 Valve 1 In-oil measurement of hydrogen and moisture Oil out Oil in Terminal blocks for RS-485 output...
  • Page 23: Status Leds

    Chapter 2 – Product overview Figure 8 Inside OPTPSU1 Power Supply Unit cabinet Circuit breaker for AC power (F1) Surge arresters Power switch (S1) Power supply Relays (3 pcs). Each relay has a LED that is lit when the relay is active. DC OK LED.
  • Page 24: Gas Level Alerts

    OPT100 User Guide M211858EN-K 2.11 Gas level alerts DGA monitor has a comprehensive selection of individual gas level alerts: • There are 4 possible alerts for each measured gas and the total dissolved combustible gases (TDCG) parameter: latest measured value, ROC 1 day, ROC 7 days, and ROC 30 days.
  • Page 25: Dnp3 Protocol

    2.13 DNP3 protocol DNP3 protocol is an optional feature and requires a license. If a license is not installed on the DGA monitor, you cannot enable the protocol. Contact Vaisala for acquiring the license. OPT100 DGA monitor provides the following functionality over DNP3 (Distributed Network Protocol): •...
  • Page 26: Iec 61850 Protocol

    2.14 IEC 61850 protocol IEC 61850 protocol is an optional feature and requires a license. If a license is not installed on the DGA monitor, you cannot enable the protocol. Contact Vaisala for acquiring the license. Data measured by DGA monitor can be read using IEC 61850 protocol over the external Ethernet connection.
  • Page 27: Autocalibration

    Chapter 2 – Product overview Protocol test mode overrides the data that is available to communication protocols with arbitrary test data. The mode is useful for verifying proper handling of data and alarms when integrating the DGA monitor with a SCADA system. The recommendation is to use the protocol test mode during commissioning, after the required output protocols have been configured and enabled.
  • Page 28: Table 9 Types Of Autocalibration

    OPT100 User Guide M211858EN-K Table 9 Types of autocalibration Autocalibration type Description Notes Initial autocalibration Initial autocalibration runs Measurement results generated automatically after a DGA during the initial monitor has been initialized autocalibration may not be and measurement started. accurate. Gas level alerts are Initial autocalibration cannot be not indicated during initial skipped, and it must complete...
  • Page 29: User Interface

    Chapter 3 – User interface 3. User interface 3.1 Overview of user interface DGA monitor runs a web server on its processing unit. This web server provides the user interface of the DGA monitor. The user interface can be accessed through Ethernet using a standard web browser;...
  • Page 30 OPT100 User Guide M211858EN-K 4. Choose the appropriate user level: • Select Admin to perform installation and maintenance tasks. • Select User to view data without making changes to the system. 5. If logging in as Admin, enter the password. The unique administration password of the DGA monitor is included in the delivery documentation but note that it can be changed by the Admin after a successful login.
  • Page 31: User Interface Basics

    Chapter 3 – User interface 3.1.2 User interface basics Figure 11 Elements of the DGA monitor user interface Current time of the DGA monitor, adjusted to the browser's time zone. DGA monitor keeps UTC time (Coordinated Universal Time) internally. Name of the currently selected user interface page Navigation for subpages (if any) Status indicator Alert indicator...
  • Page 32 OPT100 User Guide M211858EN-K 3.1.2.2 Alert indicator Alert indicator shows the presence of active alerts. The colors of the indicator change according to the severity of the most severe currently active alert: Light gray background: no active alerts. Orange background: gas caution or device warning active. Red background: gas alarm or device error active.
  • Page 33: Figure 12 Alert Summary Message Being Shown

    Chapter 3 – User interface Figure 12 Alert summary message being shown...
  • Page 34: Measurements Page

    OPT100 User Guide M211858EN-K 3.1.3 Measurements page Measurements page provides views into the measurement data in graphical and tabular formats. Gas measurements tab shows the primary graph view. Figure 13 Measurement graph Graph area. Measurement data created during autocalibration is marked with a light blue background.
  • Page 35: Figure 14 Measurement Graph For 1 Day Rate Of Change Values

    Chapter 3 – User interface Graph lines show empty gaps if data is missing from two consecutive measurement cycles or for more than three hours. Data may be missing for several reasons. For example, DGA monitor may have been turned off, in standby mode, or measurement cycles have not produced reliable results for some or all measurement parameters.
  • Page 36: Figure 15 Total Gas Pressure Graph

    OPT100 User Guide M211858EN-K You can change the pressure unit that is used to display the graph on page Settings > Units. Figure 15 Total gas pressure graph...
  • Page 37: Dga Diagnostics Page

    Customer or any other user of such data. The use and/or application of any such data shall be at the sole risk and liability of the Customer or the user. DGA Diagnostics is an optional feature and requires a license. Contact Vaisala for acquiring the license.
  • Page 38: Figure 17 Duval Triangles On The Dga Diagnostics Page

    OPT100 User Guide M211858EN-K Figure 17 Duval triangles on the DGA diagnostics page DGA Diagnostics page shows the progression of selected data points from the past year overlaid on top of Duval triangles: • Triangle 1: Classic Duval triangle for DGA in oil-filled transformers using methane, ethylene, and acetylene •...
  • Page 39: Alerts Page

    Chapter 3 – User interface The data points are 24 hour average values for the included gases and must meet the following criteria for each point: • At least 5 reliable measurement points must exist in the previous 24 hours. Reliability criteria are the same as for gas alert functionality.
  • Page 40: Figure 19 Event Details For A Gas Caution Alert

    OPT100 User Guide M211858EN-K Alerts page shows the current alert status of the DGA monitor. Top part of the page shows the number of currently active gas level alerts and device alerts by severity. More information on the individual alerts is shown in the Events section. By default, only active alerts are listed.
  • Page 41: Control

    Chapter 3 – User interface 3.1.6 Control pages DGA Monitor page shows current status of the DGA monitor and contains the controls for initialization, starting and stopping the measurement, and uninstalling the DGA monitor. Only relevant controls are shown, and they can only be used by the Admin user. Figure 20 DGA monitor Control page...
  • Page 42: Figure 21 Autocalibration Page

    OPT100 User Guide M211858EN-K Autocalibration page shows current status of autocalibration and a history of autocalibration events. Admin user can configure the automatically started autocalibration on this page and also start a manual autocalibration cycle. Figure 21 Autocalibration page...
  • Page 43: Settings

    • Select View to see the software license texts of the open source components used in the DGA monitor software. • Select Update to start the software update. To update the software version of the DGA monitor, you need a software update file from Vaisala.
  • Page 44: Changing Administrator Password

    Select Download Data to retrieve the file. The diagnostic data in the file is intended for use only by Vaisala. Figure 23 About page 3.2 Changing administrator password CAUTION! If you forget your administrator password, you need to contact Vaisala Support to reset it for you.
  • Page 45: Starting Measurement

    Chapter 3 – User interface Figure 24 Changing administrator password 1. Connect to the user interface as the Admin user. See Connecting to user interface (page 27). 2. In the top right corner of the user interface, select Admin > Change password. 3.
  • Page 46: Setting Device Name

    OPT100 User Guide M211858EN-K 1. Connect to the DGA monitor user interface and log in as an Admin user. 2. Start measurement from Control > Start Measuring. DGA monitor warms up during the first three measurement cycles. Measurements made during the warm-up cycles are not guaranteed to be within the accuracy specification.
  • Page 47: Setting Date And Time

    Chapter 3 – User interface 3.5 Setting date and time Figure 26 Date and time settings page DGA monitor uses UTC (Coordinated Universal Time) internally. Time and time stamps in the user interface are shown according to the time zone of the connecting web browser.
  • Page 48: Configuring Unit Settings

    OPT100 User Guide M211858EN-K 4. To set up time synchronization with a Network Time Protocol (NTP) server: NTP synchronization requires that the network connection is configured and the IP address of the NTP server is reachable. a. Select Use NTP. b.
  • Page 49: Configuring Autocalibration

    Figure 28 Autocalibration page Starting with software version 1.11.0, the DGA monitor performs autocalibration when needed and when environmental conditions are suitable. Vaisala recommends leaving this feature enabled, but Admin user can disable it from the Autocalibration page. Even if the automatically started autocalibration is disabled, the initial autocalibration will still run after the DGA monitor has been initialized.
  • Page 50: Configuring Gas Level Alerts

    OPT100 User Guide M211858EN-K 3. Configure the setting using the switch on the page: • If you want to disable automatically started autocalibration, move the switch to the left. • If you want DGA monitor to start autocalibration when it determines it is necessary, set the switch to position Start automatically.
  • Page 51: Muting Gas Level Alerts

    Chapter 3 – User interface 3. For each alert you want to configure: a. Change to the appropriate tab that shows the alert you want to configure. b. Enter the Caution Limit in ppm. This is the limit that triggers a gas alert with caution severity.
  • Page 52: Configuring Network Connection

    OPT100 User Guide M211858EN-K Figure 30 Alert page with methane caution alert visible 3.10 Configuring network connection Figure 31 IP address settings page...
  • Page 53: Configuring User Interface Security

    Chapter 3 – User interface This procedure configures the network settings of the external Ethernet connection. 1. Connect to the user interface as the Admin user. See Connecting to user interface (page 27). 2. In the user interface, select Settings > Network > IP Address. 3.
  • Page 54 TLS certificate (TLS 1.1 or 1.2) on the DGA monitor. However, note that HTTPS traffic is always encrypted even if the certificate is not trusted. Vaisala recommends using encrypted connections (HTTPS). 1. Connect to the user interface as the Admin user. See...
  • Page 55: Configuring Modbus Protocol

    Chapter 3 – User interface 3.12 Configuring Modbus protocol Figure 33 Modbus settings page 1. Connect to the user interface as the Admin user. See Connecting to user interface (page 27). 2. In the user interface, select Settings > Network > Modbus. 3.
  • Page 56: Configuring Dnp3 Protocol

    Figure 34 DNP3 settings page DNP3 protocol is an optional feature and requires a license. If a license is not installed on the DGA monitor, you cannot enable the protocol. Contact Vaisala for acquiring the license. 1. Connect to the user interface as the Admin user. See...
  • Page 57 Chapter 3 – User interface 4. If you enabled the DNP3 protocol, configure the following settings: Connection type Interface where DNP3 protocol will be available. Slave link address Address of this DGA monitor. Master link address Address of the DNP3 master. Default analog input variation Data format variation for analog input values.
  • Page 58: Configuring Iec 61850 Protocol

    Figure 35 IEC 61850 settings page IEC 61850 protocol is an optional feature and requires a license. If a license is not installed on the DGA monitor, you cannot enable the protocol. Contact Vaisala for acquiring the license. 1. Connect to the user interface as the Admin user. See...
  • Page 59: Using Protocol Test Mode

    Chapter 3 – User interface 3.15 Using protocol test mode Figure 36 Protocol test mode enabled 1. Connect to the user interface as the Admin user. See Connecting to user interface (page 27). 2. In the user interface, select Settings > Network > Protocol Test Mode. 3.
  • Page 60: Configuring Relays

    OPT100 User Guide M211858EN-K 5. To set your own test data: a. Edit the values of the fields as required. b. Check the checkboxes to enable the desired caution and warning level alerts. c. Select Update at the bottom of the page to take the new test data into use. 6.
  • Page 61 Chapter 3 – User interface 3. Each of the three relays is configured individually. For each relay: a. Select relay Mode: • In Normal mode, the relay is activated by the selected Trigger. • In Test mode, the relay activation is controlled manually from the Set test state buttons.
  • Page 62: Maintenance

    OPT100 User Guide M211858EN-K 4. Maintenance 4.1 Maintenance safety WARNING! Keep away from live circuits. Operating personnel must observe safety regulations at all times. WARNING! Ground the DGA monitor cabinets as instructed in the wiring instructions. Verify the grounding before and after performing maintenance on the unit.
  • Page 63: Maintenance

    These updates are only applied to new measurement cycles after the update, which may cause some discontinuity in the measured gas levels even if the concentration of gases in the measured oil has not changed. Vaisala recommends starting autocalibration manually after a software update.
  • Page 64: Figure 38 About Page With Current Software Version And Update Button

    OPT100 User Guide M211858EN-K Figure 38 About page with current software version and update button 1. Connect to the user interface as the Admin user. See Connecting to user interface (page 27). 2. Switch the DGA monitor to Standby mode. See Stopping measurement (page 64).
  • Page 65 Chapter 4 – Maintenance 6. Select Upload to send the file to the DGA monitor. 7. DGA monitor will verify that the file was received successfully, and start the update. The update may take up to 30 minutes. When the process is complate, a user interface message will show the result of the update.
  • Page 66: Starting Autocalibration Manually

    OPT100 product page http://www.vaisala.com/opt100 for updates. • If a repair procedure you need is not available in the documentation, contact Vaisala for assistance. • Some repairs can only be performed by a Vaisala representative. • For any components included inside the OPT100 DGA Monitor or the OPTPSU1 power supply unit, use only spare parts supplied by Vaisala.
  • Page 67: Turning Off The Dga Monitor For Maintenance

    Chapter 4 – Maintenance 4.6.2 Turning off the DGA monitor for maintenance 1. Switch the DGA monitor to standby mode. 2. Turn off AC power to the DGA monitor: a. Turn off AC power from the external disconnection device. b. Lock the disconnection device in the off position, if possible. c.
  • Page 68: Starting Measurement

    OPT100 User Guide M211858EN-K 2. Turn on DC power inside the power supply unit: a. Open the power supply unit cabinet. b. Turn on circuit breaker F1. c. Turn on switch S1. 3. Turn on AC power to the DGA monitor: a.
  • Page 69 Chapter 4 – Maintenance The procedure assumes that the DGA monitor is completely installed and currently measuring when the uninstallation is started. The complete procedure may take up to 3 hours. If the DGA monitor is already in standby mode, you can skip step 2 step CAUTION!
  • Page 70 OPT100 User Guide M211858EN-K 5. Push in the gray knob on the side of the bleed valve, and turn it 90° clockwise to lock the valve manually. Select Continue from the user interface when done. 6. When instructed to do so, remove the bleed plug from the bleed valve. Select Continue when done.
  • Page 71 Chapter 4 – Maintenance 8. When instructed to do so, disconnect oil pipes from the Oil In and Oil Out connections on the outside of the DGA monitor. Place the ends of the pipes in the waste oil container to collect the oil that drains out.
  • Page 72: Troubleshooting

    OPT100 User Guide M211858EN-K 5. Troubleshooting 5.1 Problem situations Table 11 Troubleshooting table Problem Possible cause Solution Green status LED is flashing. DGA monitor is in standby Start the measurement mode. mode. Starting measurement (page 43). Yellow status LED is lit. A device warning is active. Log in to the user interface.
  • Page 73: System Statuses

    HTTPS traffic is always encrypted even if the certificate is not trusted. I am unable to log in to the user Forgotten password. Contact Vaisala. See Technical interface as Admin user. support (page 101). 5.2 System statuses System status is shown in the header at the top of the user interface.
  • Page 74: Device Alerts

    • Errors typically prevent normal measurement and require user intervention to recover from. Contact Vaisala for support if you are unable to resolve a device alert yourself. From the About page of the user interface you can create a diagnostic data file that can help Vaisala diagnose the device alert.
  • Page 75: Table 13 Device Alerts

    Set new time or enable time sync in the Settings > General page. If this error reoccurs, consider changing the backup battery of the real-time clock. Contact Vaisala support for help with changing the backup battery. 2099 Measurement cycle fault...
  • Page 76: Resolving A Measurement Cycle Fault

    If the lock is open the valve must be replaced to resolve the error. Contact Vaisala support. 5.4 Resolving a measurement cycle fault Figure 39 Measurement cycle fault active Measurement cycle fault (code 2099) device alert is activated when DGA monitor has detected a problem that prevents measurement, and has automatically stopped running measurement cycles.
  • Page 77 10. If uninstalling and re-initialization did not remove the problem: a. Generate a new diagnostic data file from the DGA monitor. See About page (page 41). b. Contact Vaisala support and provide the diagnostic data file to help diagnose the error.
  • Page 78: Technical Specification

    OPT100 User Guide M211858EN-K 6. Technical specification 6.1 OPT100 technical specification Table 14 OPT100 measurement specification Parameter Range 1) 2) Accuracy Repeatability Methane (CH 0 … 10 000 ppm ±4 ppm or ±5 % of 10 ppm or 5 % of reading reading Ethane (C 0 …...
  • Page 79: Table 16 Opt100 Field Performance

    Chapter 6 – Technical specification Property Description/Value Initialization time to full accuracy Two days Data storage At least 10 years Expected operating life > 15 years Three cycles for ethane and hydrogen. Table 16 OPT100 field performance Parameter Typical variance to laboratory DGA Acetylene (C ±1 ppm or ±10 % of reading Hydrogen (H...
  • Page 80: Table 18 Opt100 Operating Environment

    OPT100 User Guide M211858EN-K Table 18 OPT100 operating environment Property Description/Value Transformer oil type Mineral oil +125 °C (+257 °F) Required minimum fire point transformer oil Transformer oil pressure at oil inlet Max. 2 bar continuous Burst pressure 20 bar Transformer oil temperature at oil inlet Max. +100 °C (+212 °F) Ambient humidity range 0 …...
  • Page 81: Table 21 Opt100 Mechanical Specifications

    Connector for service use RJ45 connector, inside OPT100 housing (temporary local use only) Table 21 OPT100 mechanical specifications Property Description/Value Oil fitting Stainless steel Swagelokâ fitting for 10 mm (0.39 in) outer diameter tubing. See list of accessories for adapters available from Vaisala.
  • Page 82: Table 22 Opt100 Type Tests

    OPT100 User Guide M211858EN-K Property Description/Value Max. length of oil pipe to transformer Max. 10 m (33 ft) with 7 mm (0.28 in) inner diameter tubing Max. 5 m (16 ft) with 4 mm (0.15 in) inner diameter tubing Recommended oil pipe properties 10 mm (0.39 in) outer diameter 1 …...
  • Page 83: Table 23 Opt100 Compliance

    Chapter 6 – Technical specification Category Standard Class/Level Test IEC61326-1 Industrial Electrical equipment for measurement, control, and laboratory use - EMC requirements FCC 47 CFR 15, section Class A Limits for conducted 15.107 emissions ISED ICES-003, section Class A Limits for conducted 5(a)(i) emissions Environmental...
  • Page 84: Opt100 Accessories And Spare Parts

    OPT100 User Guide M211858EN-K 6.2 OPT100 accessories and spare parts Table 24 OPT100 accessories Description Part number Wall Mounting Set OPTMSET1 Ground Mounting Set OPTMSET2 Weather Shield OPTSHLD1 Tubing adapter, 10 mm to 3/8" (2 pcs) ASM213275SP Tubing adapter, 10 mm to 1/4" (2 pcs) ASM213274SP Table 25 OPT100 spare parts Description...
  • Page 85: Dimensions With Ground Mounting Set

    Chapter 6 – Technical specification 6.3 Dimensions with Ground Mounting Set [in] [0.79] 510 [20.08] 675 [26.57] 453 [17.83] 550 [21.65] 400 [15.75] 350 [13.78] Figure 40 OPT100 dimensions with Ground Mounting Set...
  • Page 86: Dimensions With Wall Mounting Set

    OPT100 User Guide M211858EN-K 6.4 Dimensions with Wall Mounting Set [in] 675 [26.57] 467 [14.45] 550 [21.65] 400 [15.75] Figure 41 OPT100 dimensions with Wall Mounting Set...
  • Page 87: Figure 42 Mounting Dimensions Of Installation Beams

    Chapter 6 – Technical specification [in] Ø 8.50 holes for M8 anchors or bolts Installation beam symmetry lines Figure 42 Mounting dimensions of installation beams...
  • Page 88: Appendix A: Modbus Reference

    OPT100 User Guide M211858EN-K Appendix A. Modbus reference A.1 Default Modbus communication settings Table 26 Modbus TCP communication settings Property Specification IP address of Modbus server IP address of DGA monitor on interface ETH0 Modbus server port Table 27 Default Modbus RTU communication settings Property Specification Baud rate...
  • Page 89: Modbus Registers

    Appendix A – Modbus reference A complete 32-bit floating point or 32-bit integer value should be read and written in a single Modbus transaction. CAUTION! Reading the measurement data registers with incorrect floating point format setting may occasionally result in correct-looking, but nevertheless incorrect values.
  • Page 90 OPT100 User Guide M211858EN-K Register number Address Register Data format Unit and notes description 001C Carbon monoxide 32-bit float 001D 001E Carbon dioxide 32-bit float 001F 0020 Total dissolved 32-bit float combustible 0021 gases (TDCG) 0022 Hydrogen 32-bit float 0023 0024 Moisture in oil 32-bit float...
  • Page 91 Appendix A – Modbus reference Register number Address Register Data format Unit and notes description 003A Carbon monoxide 32-bit float 003B 003C Carbon dioxide 32-bit float 003D 003E Total dissolved 32-bit float combustible 003F gases (TDCG) 0040 Hydrogen 32-bit float 0041 0042 Moisture in oil...
  • Page 92 OPT100 User Guide M211858EN-K Register number Address Register Data format Unit and notes description 005E Hydrogen 32-bit float 005F Rate of change (ROC), 7 days 0064 Methane 32-bit float 0065 0066 Acetylene 32-bit float 0067 0068 Ethylene 32-bit float 0069 006A Ethane 32-bit float...
  • Page 93 Appendix A – Modbus reference Register number Address Register Data format Unit and notes description 0080 Carbon monoxide 32-bit float 0081 0082 Carbon dioxide 32-bit float 0083 0084 Total dissolved 32-bit float combustible 0085 gases (TDCG) 0086 Hydrogen 32-bit float 0087 Gas ratios (24 hour average values) 008C...
  • Page 94: Alert Registers

    OPT100 User Guide M211858EN-K A.4.2 Alert registers Gas level alert registers are 16-bit fields. Each register communicates the status of alerts of specific severity for a particular timeframe (for example, gas level caution alerts for present values). If an alert is active, the corresponding bit is set to 1. The bit is set to 0 when the alert is deactivated, turned off, or gas level alerts are muted.
  • Page 95 Appendix A – Modbus reference Register number Address Register Data format Bit order description 000C Gas level caution 16-bit field 0 = Methane alerts for 7 day 1 = Acetylene average ROC 2 = Ethylene values 3 = Ethane 4 = Carbon monoxide 5 = Carbon dioxide...
  • Page 96 OPT100 User Guide M211858EN-K Register number Address Register Data format Bit order description 000F Gas level alarms 16-bit field 0 = Methane for 24 hour 1 = Acetylene average ROC 2 = Ethylene values 3 = Ethane 4 = Carbon monoxide 5 = Carbon dioxide...
  • Page 97: Status Registers

    Appendix A – Modbus reference A.4.3 Status registers Only the first bit of each status bit field is used. Table 31 Status registers (read-only) Register Address Register Data format Notes number description 0000 Device stopped 16-bit field 0 = Device measuring normally 1 = Device stopped (either automatically or by administrator), being initialized, or being uninstalled...
  • Page 98: Exception Responses

    OPT100 User Guide M211858EN-K Register number Address Data format Test value 00D2 64-bit integer -1234567890123456789 00D3 00D4 00D5 00D6 32-bit float 1.23456 00D7 00D8 64-bit float 1.23456789 00D9 00DA 00DB A.5 Exception responses Table 33 Modbus exception responses Code Name Reason ILLEGAL FUNCTION Unsupported function code ILLEGAL DATA ADDRESS Register address or Object ID out of valid ranges ILLEGAL DATA VALUE...
  • Page 99: Modbus Communication Examples

    Appendix A – Modbus reference A.6 Modbus communication examples Reading acetylene value Device address used in the following examples is 240 (F0 The values returned by the device differ depending on the measurement conditions. Your device may not return the same values. Request Bytes on the line (hexadecimal) Description...
  • Page 100 OPT100 User Guide M211858EN-K Response Modbus RTU checksum (silence for > 3.5 bytes) End of Modbus RTU frame Communication description Register number 23 (1-based Modbus register number) = address 0016 (0-based address used in actual communication). Data format Two 16-bit Modbus registers interpreted as IEEE 754 binary32 floating point value, least significant word first.
  • Page 101: Index

    Index Index accessories...............82 IEC 61850..............24 alert indicator............30 IEC 61850 protocol autocalibration..........25, 47 configuring............56 starting..............64 maintenance..............calculated parameters...........10 remote access............61 calibration..............software update............61 cleaning..............60 turning off..............65 turning on.............. 65 measured parameters........... 10 date and time............45 measurement............32 device alerts.............72 starting............
  • Page 102 OPT100 User Guide M211858EN-K connectors............18, 19 settings pages............41 ground mounting set...........14 interior..............20 viewing data............32 wall mounting set..........16 password..............42 protocol test mode......... 24, 57 relative saturation........... 12 relays................. 58 repair................. 64 safety..............65 safety..............60, 65 serial line..............53 software update............
  • Page 103: Maintenance And Calibration Services

    Maintenance and calibration services Vaisala offers comprehensive customer care throughout the life cycle of our measurement instruments and systems. Our factory services are provided worldwide with fast deliveries. For more information, see www.vaisala.com/ calibration. • Vaisala Online Store at store.vaisala.com is available for most countries.
  • Page 106 www. v aisala.com...

Table of Contents