Cross Multiplexer Testing - HP 37717C Operating Manual

Communications performance analyzer, pdh / dsn
Hide thumbs Also See for 37717C:
Table of Contents

Advertisement

PDH / DSn Testing With The HP 37717C

Cross Multiplexer Testing

Cross Multiplexer Testing

Options Required Structured PDH UKJ (USA)or UKN (USE)
Application
For comprehensive testing of network equipment it is essential that the test
equipment can multiplex/demultiplex the test signal.
The insertion of tributary signals into the PDH multiplexer, which are then
multiplexed into the 140 Mb/s PDH structure should take place without introducing
errors. The insertion and structuring process is tested by adding a test pattern to the
tributary inserted at the tributary insert port. At the high rate side of the PDH
multiplexer the tributary is destructured and a BER test performed.
By using a protected monitor point at the high rate side of the PDH multiplexer the
mux/demux need not be taken out of service.
Default (Known State) Settings
It is advisable to set the HP 37717C to a known state before setting up a
measurement. This clears all previous settings and provides a clearly defined
instrument state. For a list of Default Settings and the procedure for accessing them
see Stored Settings in the Mainframe Operating Manua.
Cross Multiplexer Testing Test Setup Procedure
In this setup a 2 Mb/s Framed tributary, containing a test pattern, is inserted at the
tributary insert port of the PDH multiplexer. The 140 Mb/s structured PDH signal is
obtained from a protected monitor point. The 2 Mb/s tributary is destructured by the
HP 37717C test set and an Error measurement is performed on the 2 Mb/s tributary
test pattern.
A SINGLE test period of 24 HOURS is used and the printer is enabled to record
results and alarms.
The HP 37717C PDH/SDH test set GRAPHICS function is enabled. The graphical
results can be viewed on the
GRAPH
display.
38