5.1.1
Show Log Data
SHOW LOG refers to the 1-second data that is logged during a LOG key
press, an Auto Log function, or from a Trigger (User, Power Quality, or
from the optional Auxiliary I/O Module).
5.1.2
Show User Event Records
SHOW USER refers to the event record, including the exact time and date
of the User programmed event. It does not refer to the data that may be
stored before, during and after the event.
5.1.3
Show Power Quality Event Records
SHOW PWR QUAL refers to the event record, including the exact time
and date of the power-quality event. It does not refer to the data that
may be stored before, during and after the event.
5.1.4
Erase Log Data
Press the ERASE LOG DATA soft key to clear all the data from the flash
memory card.
The following pages describe, in short steps and in more detail, how to
view and download data from the 933A.
5.2
Steps to Locate & Review Records
Listed below are steps to assist you with identifying and locating the
specific data you want to review. Basically, there are two types of records
stored in the 933A: data records and event records. Decide which record
type you wish to view; a record of the event, or the data recorded during
the event?
5.2.1
Locating the Data Logged During an Event
To find actual data measured by the 933A, proceed as follows:
1. Press the keys, REVIEW > SHOW LOG > REVIEW SELECTIONS
2. Press the appropriate soft key that indicates the source of the data
- Event Logs, Key Logs, Auto Logs or All Logs.