Measuring channel calibration (after sensor exchange)

1
When replacing the sensor you should proceed in concordance with chapter 4.3 "Mounting
of sensors".
2
Press the MENU key, switch to "change" with the
ENTER key.
3
Enter the password (if active)
4
Switch to "adjustment parameter" and confirm by pressing the ENTER key.
5
At "measuring channel calibration" choose "yes" and confirm by pressing the ENTER key.
6
Set the sheet thickness for the measuring channel calibration and confirm by pressing the
ENTER key.
7
Place the calibration sheet between the sensors and confirm by pressing the ENTER key.
8
Confirm the completion of the measuring channel calibration by pressing the ENTER key.
9
Return into the measurement mode by pressing the MENU key