Troubleshooting - Dentsply Sirona X-Smart Plus User Manual

Table of Contents

Advertisement

Troubleshooting

9
TROUBLESHOOTING
When trouble is found, check the following points before contacting your
distributor. If none of these are applicable or the trouble is not remedied
even after action has been taken, the product may have failed. Contact
your distributor.
®
X-Smart
Plus control unit and AC adapter
Problem
The power is not turned
on.
AC-IN lamp does not
light.
The AC adapter does
not work.The CHRG
lamp does not light.
50/60
Cause
The AC adapter is not
connected.
The plug of the AC adapter is
not inserted into the outlet, or
there is no electricity in the
outlet.
The battery is flat.
No battery is inserted.
The internal fuse has burnt.
The AC adapter is not
connected.
The plug of the AC adapter is
not inserted into the outlet, or
there is no electricity in the
outlet.
The internal fuse has burnt.
The AC adapter fuse has
burnt.
No battery is inserted.
The battery is fully charged
or in a state near full charge.
Solution
Check the connection.
Check the connection.
Charge the battery or use
the AC adapter.
Insert battery, or use the
AC adapter.
Contact your distributor.
Check the connection.
Check the connection.
Contact your distributor.
Contact your distributor.
Insert battery.
No problem.