Auto Reverse Function - Dentsply Sirona X-Smart Plus User Manual

Table of Contents

Advertisement

Step-by-Step Instructions
6.6.4

Auto Reverse Function

There are 3 different auto reverse modes:
AUTO REVERSING:
If, during operation the load reaches the preset torque limit value, the
motor handpiece will automatically rotate in the reverse direction. When
the load is removed, the motor handpiece returns to normal forward
rotation automatically.
Load within the Torque Limit
Value
Forward Rotation
AUTO STOP:
If, during operation the load reaches the preset torque limit value, the
motor handpiece will automatically rotate in the reverse direction. When
the load is removed, the motor handpiece stops. The LCD panel shows
"
forward again, press the ON/OFF key twice.
Load within the Torque Limit
Value
Forward Rotation
BENXSPSDFUWEB / Rev.09 / 10-2018
" and the rotation speed alternately.If you want the file to rotate
Further Loading
beyond the preset 
Torque Limit Value
Reverse Rotation
Fig. 8
Further Loading
beyond the preset
Torque Limit Value
Reverse Rotation
Fig. 9
If the load is kept applied,
the file will rotate in reverse.
When the load is removed,···
Forward Rotation
If the load is kept applied, the
file will rotate in reverse. When
the load is removed,···
STOP