Dentsply Sirona X-Smart Plus User Manual page 9

Table of Contents

Advertisement

Warnings
The plastic enclosure is not sealed, do not use any liquid or spray
directly on the unit, especially on the monitor or near the electrical
sockets.
Do not crimp the cords exiting the motor handpiece and 
AC adapter.
This device is equipped with an electronic circuit to limit file
separation. Nevertheless, files may still separate due to incorrectly
set torque or speed, or the use of fatigued files.
Follow the file manufacturer's instructions for use of the endodontic
files.
The file system shown on the display must always match the file in
use. This is of the utmost importance in order to avoid misusing
reciprocating files and continuous rotary files.
Do not use files designed for continuous rotation in reciprocating
motion.
Do not use files designed for reciprocating motion in continuous
rotation.
Torque and speed values are subject to change by the file
manufacturers without notice. Therefore, the preset values in the
library must be checked prior to use. Torque values shown on the
screen are accurate and reliable only with X-Smart
contra-angles properly maintained and lubricated.
The accuracy of movement provided by the motor is guaranteed
only if the original X-Smart
properly maintained and lubricated (for more details see chapter 
6.11.2 Lubricating the
Do not use any other contra-angle or other reduction rate other than
the original one.
Do not insert any file in the contra-angle during calibration.
Calibrate every time the contra-angle is lubricated or replaced after
sterilizing, or at least once a week (see chapter
When lubricating the contra-angle, carefully check that no lubricant
penetrates the motor handpiece.
BENXSPSDFUWEB / Rev.09 / 10-2018
®
Plus 6:1 contra-angle is used,
Contra-angle).
®
Plus 6:1
6.5.6
Calibration).
9/60