Before changing the contra-angle or file, turn off the power of the
unit. Changing with the power kept on may cause unintended
rotation by accidental touch of the ON/OFF key.
Always clean the shank of the file to be installed. Allowing dirt to
enter the chuck could cause loss of concentricity and deterioration
of chucking force.
Pay attention to the direction of the battery connector when
installing. Forcible setting in the wrong direction may cause damage
and fluid leakage due to a short circuit.
Fully-charged rechargeable batteries generally discharge gradually
over time even though the device is not used. It is recommended to
recharge the battery just before use.
Should the device stop automatically due to low battery voltage, it
may not indicate low voltage immediately when the device is
powered on again.
Recharge rechargeable battery after it depletes as much as
possible. Repeating short-time use and subsequent recharging may
shorten its operating time due to the "memory effect." The battery
may recover after repeating complete discharge and full charge for
a few times (see chapter
The used Nickel Metal Hydride batteries are recyclable, but their
disposal may sometimes not be permitted by law in your location.
Return them to your distributor.
When disposing of the control unit, follow the instructions of your
local government for disposal, as they contain materials which may
become industrial waste.
When discarding the contra-angle and motor handpiece, dispose of
them as medical waste.
This product does not consider a patient's age, gender, weight or
nationality.
No special training is required for this device.
The manufacturer declines any responsibility in the case of:
Use of the device for applications other than those specified in the
instructions for use and maintenance.
BENXSPSDFUWEB / Rev.09 / 10-2018
6.9 Battery
Refresh).