Precautions
•
Before changing the contra-angle or file, turn off the power of the
unit. Changing with the power kept on may cause unintended
rotation by accidental touch of the ON/OFF key.
•
Always clean the shank of the file to be installed. Allowing dirt to
enter the chuck could cause loss of concentricity and deterioration
of chucking force.
•
Pay attention to the direction of the battery connector when
installing. Forcible setting in the wrong direction may cause damage
and fluid leakage due to a short circuit.
•
Fully-charged rechargeable batteries generally discharge gradually
over time even though the device is not used. It is recommended to
recharge the battery just before use.
•
Should the device stop automatically due to low battery voltage, it
may not indicate low voltage immediately when the device is
powered on again.
•
Recharge rechargeable battery after it depletes as much as
possible. Repeating short-time use and subsequent recharging may
shorten its operating time due to the "memory effect." The battery
may recover after repeating complete discharge and full charge for
a few times (see chapter
•
The used Nickel Metal Hydride batteries are recyclable, but their
disposal may sometimes not be permitted by law in your location.
Return them to your distributor.
•
When disposing of the control unit, follow the instructions of your
local government for disposal, as they contain materials which may
become industrial waste.
•
When discarding the contra-angle and motor handpiece, dispose of
them as medical waste.
•
This product does not consider a patient's age, gender, weight or
nationality.
•
No special training is required for this device.
The manufacturer declines any responsibility in the case of:
•
Use of the device for applications other than those specified in the
instructions for use and maintenance.
BENXSPSDFUWEB / Rev.09 / 10-2018
6.9 Battery
Refresh).
11/60