Should you at any time wish to stop the calibration process, turn
the power off.
Calibrate every time the contra-angle is lubricated or replaced
after sterilizing, or at least once a week (see chapters
Lubricating the Contra-angle
and
This function does not operate if the AC adapter is disconnected.
Do not touch or apply a load to the contra-angle chuck during
calibration.
6.5.7

Sound Volume Adjustment

The sound volume can be adjusted to 3 different levels: high, low and
limited off (the sound appears with a low level at the time of confirmation
and error, but it does not sound during reverse rotation, nor when
reaching the torque limit value..
Press the sound volume (12) key (see
1.
The sound volume and the volume mark on the LCD panel change
2.
NOTE
The last sound volume is retained, even if the power is turned off.
If "reset to default parameters" is carried out, the sound is set to
the high level.