Warnings
In order to avoid possible risks due to electromagnetic interference,
do not use any electrical medical device or electrical device of any
other kind in close proximity to the X-Smart
electromagnetic radiation emitted by the device is below the
recommended limits set forth in pertinent regulations in force
(EN 60601-1-2).
Do not use the device in the presence of free oxygen or anesthetic
substances or flammable products.
The device may possibly malfunction if used in the presence of an
electromagnetic interference wave. Do not install the 
X-Smart
waves.
The use of accessories, transducers and cables other than those
specified, with the exception of transducers and cables sold by the
manufacturer of the X-Smart
internal components, may result in increased emissions or
decreased immunity of the X-Smart
X-Smart
other equipment. In the instance that if adjacent or stacked use is
necessary, the X-Smart
operation in the configuration in which it will be used.
None of the X-Smart
or sterilized: components such as control unit, micromotor and
micromotor cable need to be disinfected, the contra-angle needs to
be sterilized prior to first use and in between each patient!
Never place the motor handpiece or any other device accessories
in an autoclave unit or ultrasonic tank.
None of the components of the X-Smart
(contra-angle excluded, see chapter
and
Do not immerse the device in ultrasonic cleaners.
8/60
®
Plus in the vicinity of any device that emits magnetic
®
Plus should not be used adjacent to or stacked with
®
Sterilization).
®
Plus as replacement parts for
®
Plus.
®
Plus should be observed to verify normal
Plus components are delivered disinfected
6.12 Cleaning, Disinfection
BENXSPSDFUWEB / Rev.09 / 10-2018
®
Plus. The
®
Plus may be sterilized