Error Code
Error
code
E-10
E-11
At the
E-12
time of
charging
E-13
E-14
E-15
Other
E-16
E-18
At the
time of
calibration
E-19
BENXSPSDFUWEB / Rev.09 / 10-2018
Error
Cause
The battery current is
too low or too high.
Battery current
The battery is flat or
not inserted.
Display
Faulty display driver.
The battery voltage is
too low. The battery is
Low battery
not inserted or the
voltage
battery operating life
is over.
The battery voltage is
High battery
too high (circuit
voltage
malfunction).
Beyond the range of
Beyond the
working temperature
range of
or damage to the
working
thermistor in the
temperature
battery section.
Abnormal heat
The battery
generation
generates abnormal
from battery
heat.
LCD panel
Faulty LCD.
Beyond the
The operating life of
upper limit
the motor handpiece
or contra-angle has
Below the
expired.
lower limit
Check/Remedy
Put the battery into
the battery
compartment or
replace the battery.
Contact your
distributor.
Put the battery into
the battery
compartment or
replace the battery.
Contact your
distributor
Use within the range
of working
temperature or
replace the battery.
Replace the battery.
If abnormal heat
generates from the
new battery as well,
malfunction of the
circuit may be
suspected. Contact
your distributor.
Contact your
distributor.
Replace the motor
handpiece or contra-
angle.