Dentsply Sirona X-Smart Plus User Manual page 42

Table of Contents

Advertisement

Step-by-Step Instructions
5
Sterilization
6
Storage
42/60
Steam sterilization at
134°C (274°F) for
3 minutes.
Keep the contra-
angle in sterilization
packaging in a dry
and clean
environment.
Use only autoclaves that are
matching the requirements of
EN 13060, EN 285.
Use a validated sterilization
procedure according to 
ISO 17665.
Respect the maintenance
procedure of the autoclave
device given by the
manufacturer.
Use only this recommended
sterilization procedure.
Control the efficiency
(packaging integrity, no
humidity, color change of
sterilization indicators,
physico-chemical integrators,
digital records of cycles
parameters).
Maintain traceability of
procedure records.
Sterility cannot be guaranteed
if packaging is open, damaged
or wet.
Check the packaging and the
contra- angle before using it
(packaging integrity, no
humidity and validity period).
BENXSPSDFUWEB / Rev.09 / 10-2018