Step-by-Step Instructions
5
Sterilization
6
Storage
42/60
Steam sterilization at
134°C (274°F) for
3 minutes.
Keep the contra-
angle in sterilization
packaging in a dry
and clean
environment.
•
Use only autoclaves that are
matching the requirements of
EN 13060, EN 285.
•
Use a validated sterilization
procedure according to
ISO 17665.
•
Respect the maintenance
procedure of the autoclave
device given by the
manufacturer.
•
Use only this recommended
sterilization procedure.
•
Control the efficiency
(packaging integrity, no
humidity, color change of
sterilization indicators,
physico-chemical integrators,
digital records of cycles
parameters).
•
Maintain traceability of
procedure records.
•
Sterility cannot be guaranteed
if packaging is open, damaged
or wet.
•
Check the packaging and the
contra- angle before using it
(packaging integrity, no
humidity and validity period).
BENXSPSDFUWEB / Rev.09 / 10-2018