Recommended separation distances between portable and mobile RF
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the device as
recommended below, according to the maximum output power of the communications
equipment.
Rated
maximum
output power
of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
BENXSPSDFUWEB / Rev.09 / 10-2018
communications equipment and the device
Separation distance according to frequency of transmitter
150 kHz to
80 MHz
d = 1,2 √P
0.12
0.38
1.2
3.8
12
(m)
80 MHz to
800 MHz
d = 1,2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2,5 GHz
d = 2,3 √P
0.23
0.73
2.3
7.3
23