Step-by-Step Instructions
NOTE
There is no need to turn the power on to charge the battery.
The AC-IN lamp lights to indicate that the supply power is on. It
does not go out even after a charge is completed. See the CHRG
lamp to check the charging conditions.
The standard charging time is about 5 hours, but it varies
depending on use, condition of the battery, temperature and
whether the battery is new or old. Older batteries may remarkably
shorten both the charging and operating time.
When charging, the temperature of the battery is measured.
Therefore, if this product is placed in an environment where the
temperature changes rapidly (for example, near to a window
exposed to sunlight, near the outlet of an air conditioner, or
heater), the battery cannot be properly charged. Store and charge
this product in a place with low variations in temperature.
The battery automatically charges when connected to the mains,
even when the power is on. But when the motor handpiece is in
use, charging is suspended to protect the battery.
In the following cases, it may not start charging:
- The temperature of the battery is too low or too high (less than
0°C (32°F) or more than 40°C (104°F)).
- The voltage of the battery is sufficient (this does not necessarily
mean fully charged).
- Battery is not connected.
- The voltage of the battery is abnormal (see chapter
Code).
During approximately 3 seconds after the connection or
disconnection of the AC adapter, the control unit cannot be
switched on.