Step-by-Step Instructions
6.5.6

Calibration

This function is to decrease fluctuation in the rotation speed of the motor
handpiece and the difference in torque by the contra-angle.
Calibration is recommended when using a new/other contra-angle or
after an extended period of operation, as the running properties can
change with usage, cleaning and sterilization.
Turn the power off.
1.
2.
Attach the X-Smart
Connect the AC adapter and check that the AC-IN lamp lights.
3.
Turn the power on.
4.
Hold down the CAL (10) key (see
5.
During the calibration process the display will read:
The motor handpiece begins to rotate: leave it as it is until it stops.
When the calibration process is completed, the rotation stops and
the display reads:
24/60
®
Plus 6:1 contra-angle to the motor handpiece.
Fig.
1) for more than 2 seconds.