Adjustment Of The Start Time (Casting Time Function) - Dentsply Sirona inFire HTC speed Operating Instructions Manual

Hide thumbs Also See for inFire HTC speed:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions inFire HTC speed
Explanation
The correction program
Call
61 30 459 D 3497
D 3497.201.01.19.02 09.2017
7.2
Adjustment of the start time (casting time
function)
When calculating the start time (if the casting-time function is used), the
furnace controller assumes an average voltage of 230 V.
The local voltage may significantly differ from this value according to local
circumstances.
Accordingly, the calculation of the start and end time of the controller may
be incorrect.
Substantial variations (of more than 30 minutes) may be adjusted by
means of the correction program.
The correction program compares the values in a test run and
automatically corrects them.
This correction also remains intact even after the furnace is switched off.
NOTICE
Furnace empty?
No sintering trays may be located in the furnace during the correction
program run.
1. Load program 00.
2. Start the program.
 The program automatically switches off following the correction
measurement.
NOTICE
Long program run
This program takes about 3-4 hours to run.
➢ You can quit the program by pressing the START/STOP key if
necessary.
7.2 Adjustment of the start time (casting time function)
7 Default settings
45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents