Dsp Not Running; No Run Data Available In Device - Agilent Technologies 1260 Infinity G1315C User Manual

Diode array and multiple wavelength detector
Table of Contents

Advertisement

DSP Not Running

This error message comes up when the communication between the optical
unit and the main board has a problem.
Probable cause
1
2
3

No Run Data Available In Device

In a very rare case the capacity of the CompactFlash Card is not sufficient.
This could happen for example when the interrupt of LAN communication
takes longer and the detector uses special settings (e.g full data rate at 80 Hz
plus full spectra plus all signals) during data buffering.
Probable cause
1
Agilent 1260 Infinity DAD and MWD User Manual
Random communication error.
Defective detector main board.
Defective PDA/optical unit.
CompactFlash Card is full.
Error Information
Detector Error Messages
Suggested actions
• Switch the detector off and on again at the
power switch. If the error reoccurs:
• Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
• Correct communication problem.
• Reduce data rate.
7
141

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents