Table of Contents

Advertisement

OdaLog® L2 – Gas Logger User Manual
 Carefully remove the filter housing assembly.
 Remove and retain the O-ring.
 Remove filter.
 Remove, clean, and replace stainless filter grill (if fitted)
NOTE: Ensure hands are free of dirt and grease prior to handling new filter.
 Replace filter with "shiny" side facing outwards.
 Replace O-ring. DO NOT GREASE O-RING
 Replace filter housing assembly on OdaLog® body (aligning screw holes)
 Replace three end cap-retaining screws. DO NOT OVERTIGHTEN
Note 1: Filter orientation is important. Ensure the replacement filter is correctly placed with the "shiny"
side facing outward.
Note 2: Some sensor filters are fitted with stainless wire gauze protection. If fitted, this should be cleaned
and dried before fitting a new filter.

8.2. Service

To locate your nearest Thermo Fisher Scientific service centre please contact us or refer to our website.
P/N: 10-1000
Revision G, January 2024
Page | 21

Advertisement

Table of Contents
loading

Table of Contents