OdaLog® L2 – Gas Logger User Manual
Carefully remove the filter housing assembly.
Remove and retain the O-ring.
Remove filter.
Remove, clean, and replace stainless filter grill (if fitted)
NOTE: Ensure hands are free of dirt and grease prior to handling new filter.
Replace filter with "shiny" side facing outwards.
Replace O-ring. DO NOT GREASE O-RING
Replace filter housing assembly on OdaLog® body (aligning screw holes)
Replace three end cap-retaining screws. DO NOT OVERTIGHTEN
Note 1: Filter orientation is important. Ensure the replacement filter is correctly placed with the "shiny"
side facing outward.
Note 2: Some sensor filters are fitted with stainless wire gauze protection. If fitted, this should be cleaned
and dried before fitting a new filter.
8.2. Service
To locate your nearest Thermo Fisher Scientific service centre please contact us or refer to our website.