Configure process measurement
Table 4-3:
Flow Direction setting
Negate Bidirectional
(1) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Effect of flow direction on flow totals
Flow direction affects how flow totals and inventories are calculated.
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
4.5

Configure density measurement

The density measurement parameters control how density is measured and reported.
4.5.1

Configure Density Measurement Unit

Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit controls the measurement units that will be used in density
calculations and reporting.
Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
42
Effect of the flow direction on flow values
Forward
Totals increase
Totals do not change
Totals increase
Totals increase
Totals do not change
Totals decrease
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS
Device Tools > Configuration > Process Measurement > Density
Micro Motion Model 2700 Transmitters with Intrinsically Safe Outputs
(continued)
Actual flow direction
Forward
Zero flow
Negative
0
Actual flow direction
Zero flow
Totals do not change
Totals do not change
Totals do not change
Totals do not change
Totals do not change
Totals do not change
Reverse
Positive
Reverse
Totals do not change
Totals increase
Totals decrease
Totals increase
Totals increase
Totals increase