Start Level (Start Lev.); Stop Level (Stop Level); Zero Clear At The Start Of Hold Section (Start Clr); Sample Hold - Unipulse TM400 Operation Manual

Portable torque monitor
Table of Contents

Advertisement

Start level (START LEV.)

Sets the level to start detection.
<Setting range>

Stop level (STOP LEVEL)

Sets the level to stop detection.
<Setting range>

Zero clear at the start of hold section (START CLR)

Set whether or not to reset the angle to zero at the start of hold section.
* This function works only when the peak hold (angle) is selected.
<Setting range>
- OFF:
- ON:
* Do not use this function in combination with "zero clear from the recording start (REC.ST.CLR)"
that is described in encoder setting.
6-2.

Sample hold

Holds an arbitrary point.
- Operation of sample hold
key input
HOLD
HOLD display
Holds the indicated value at
Press the
27
-9999 to 9999
-9999 to 9999
OFF, ON
The angle is not reset to zero at the start of hold section.
The angle is reset to zero at the start of hold section.
Lights off
HOLD
key down.
HOLD
key again to release hold.
Hold section
Lights on
Lights off
6 Hold Function
Chapter
6
Torque
Indicated value
Time
27