Preparing for Use

Power Requirements

Power Requirements
The analyzer is a portable instrument and requires no physical
installation other than connection to a power source. There is no need
to select a line voltage.
Table 1-2
AC Power Requirements
Voltage
Voltage
Power Consumption, On
Power Consumption, Standby
Table 1-3
DC Power Requirements
Voltage
Power Consumption

Checking the Fuse

Where IEC regulations apply, use a 5 by 20 mm, rated F5A, 250 V IEC
approved fuse. This fuse may be used with input line voltages of 115 V
or 230 V. Its part number is 2110-0709.
Where UL/CSA regulations apply, use a 5 by 20 mm rated fast blow,
5 A, 125 V UL/CSA approved fuse (part number 2110-0756). This fuse
may only be used with an input line voltage of 115 V.
The line fuse is housed in a fuse holder in the upper left hand corner of
the rear panel.
To remove the fuse, first disconnect the power cord from the
instrument. Then insert the tip of a screwdriver into the slot at the
middle of the fuse holder, and turn counterclockwise to extend the fuse
holder.
For continued protection against fire hazard, replace the line
WARNING
fuse only with the same type and rating. The use of other fuses
or material is prohibited.
1-4
90-132 Vrms (47 to 440 Hz)
195 - 250 Vrms (47 to 66 Hz)
less than 300 W
less than 5 W
12-20 Vdc
less than 200 W