Analyzer Accuracy and the Internal Alignment Process

Analyzer Accuracy and the Internal Alignment
Process
Data from the internal alignment routine is necessary for spectrum
analyzer operation. The internal alignment routine runs continuously
to ensure that the analyzer is using current alignment data that
improves the analyzer frequency and amplitude accuracy. Press the
,
keys to view the alignment menus. For more
System
Alignments
detailed information on the alignment keys, refer to
Chapter
6.

Warm-up Time

In order for the analyzer to meet its specifications, allow it to warm up
for 5 minutes after being turned on before attempting to make any
calibrated measurements. If the analyzer is an Agilent E4402B,
E4403B, E4404B, E4405B, E4407B, or E4408B, connect a BNC cable
between AMPTD REF OUT and INPUT 50
5 minute warm-up, press
is selected, refer to the analyzer specifications for the
Align Off
conditions required to maintain calibration.
2-24
using an adapter. After a
,
,
System
Alignments
Align Now
"Alignments" in
,
. If
All
Auto
Chapter 2