Procedure (6.01 Ghz To 12.0 Ghz) E4405B, E4407B, E4408B - Agilent Technologies ESA-E Series User Manual

Esa spectrum analyzers, 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Functional Testing
Displayed Average Noise Level
5. Press
. Set the center frequency of the analyzer to the
Center Freq
frequency recorded in the Measured Frequency column (h) in
3-1
for 6.01 GHz to 6.7 GHz.
6. Press the following keys on the analyzer:
,
BW/Avg
1 kHz
Video BW Auto Man
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
,
Display
Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 11 in
3-1. The Average Noise Level should be less than the Maximum.
Procedure (6.01 GHz to 12.0 GHz) E4405B, E4407B,
E4408B
1. Press
,
System
Power On/Preset
analyzer. Wait for the preset routine to finish. Set the analyzer by
pressing the following keys:
,
FREQUENCY
Start Freq
,
Stop Freq
12.0 GHz
,
AMPLITUDE
–70 dBm
,
Attenuation (Man)
,
BW/Avg
1 MHz
Video BW Auto Man
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average (On)
Single
Wait until AVG 3 is displayed to the left of the graticule (the
analyzer will take 3 sweeps and then stop).
3. Press
Peak Search
column (i) in
Table 3-1
3-16
,
30 Hz
,
Preset (Factory)
,
6.01 GHz
0 dB
,
10 kHz
,
3, Enter
. Record the frequency in the Measured Frequency
for 6.01 GHz to 12.0 GHz.
Table
Table
,
on the
Preset
Chapter 3