Procedure

1. Connect the equipment as shown in
2. Set the synthesized sweeper controls as follows:
75 Ω Input only. Connect the minimum loss adapter between the low
NOTE
pass filter and 75 Ω Input.
3. Press
4. Adjust the synthesized sweeper amplitude to place the peak of the
5. Set the analyzer by pressing the following keys:
6. Wait for two sweeps to finish, then press the following analyzer keys:
7. Press the ⇑ key on the analyzer to step to the second harmonic (at
Chapter 3
,
Frequency
40 MHz
(50 Ω Input only)
,
Amplitude
–10 dBm
(75 Ω Input only)
,
Amplitude
–4.3 dBm
,
System
Power On/Preset
analyzer. Wait for the preset routine to finish.
Set the analyzer by pressing the following keys:
,
FREQUENCY
40 MHz
,
SPAN
1 MHz
,
AMPLITUDE
–10 dBm
,
AMPLITUDE
44 dBmV
,
Attenuation (Man)
10 dB
,
BW/Avg
30 kHz
signal at the reference level.
,
SPAN
50 kHz
,
BW/Avg
1 kHz
,
Video BW Auto Man
100 Hz
Peak Search
Mkr →
Mkr → CF Step
,
Marker
Delta
FREQUENCY
80 MHz). Press
Peak Search
should be less than the Maximum value listed in the Test Limits
Table.
Second Harmonic Spurious Responses
Figure
,
Preset (Factory)
(50 Ω Input only)
(75 Ω Input only)
. The marker delta amplitude reading
Functional Testing
3-6.
,
on the
Preset
3-45