Functional Testing
Displayed Average Noise Level
7. Press the following keys on the analyzer:
,
Display
Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 14 in
Table
3-1. The Average Noise Level should be less than the
Maximum.
Procedure (22.01 GHz to 26.5 GHz) E4407B and
E4408B
1. Press
,
System
Power On/Preset
analyzer. Wait for the preset routine to finish.
Set the analyzer by pressing the following keys:
,
FREQUENCY
Start Freq
,
Stop Freq
26.5 GHz
,
AMPLITUDE
–70 dBm
,
Attenuation (Man)
,
BW/Avg
1 MHz
Video BW Auto Man
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average (On)
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer
will take 3 sweeps and then stop).
3. Press
Peak Search
column (l) in
Table 3-1
3-20
,
Preset (Factory)
,
22.01 GHz
0 dB
,
10 kHz
,
,
3
Enter
. Record the frequency in the Measured Frequency
for 22.01 GHz to 26.5 GHz.
,
on the
Preset
Chapter 3