Procedure - Agilent Technologies ESA-E Series User Manual

Esa spectrum analyzers, 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Instrument
Additional Equipment for 75 Ω Input
Pad, minimum loss
Type-N(f), to
BNC (m)
Figure 3-5

Procedure

1. Connect the equipment as shown in
2. Preset the synthesized sweeper. Set the synthesized sweeper
3. Press
Chapter 3
Critical Specifications
(for this test)
controls as follows:
,
CW
50 MHz
(50 Ω Input)
,
Power Level
0 dBm
(75 Ω Input)
,
Power Level
4 dBm
,
System
Power On/Preset
analyzer, then wait for the preset routine to finish. Press
,
Alignments
Auto Align
keys:
,
FREQUENCY
50 MHz
,
SPAN
45 kHz
,
AMPLITUDE
Attenuation Auto Man
,
BW/Avg
Resolution BW Auto Man
,
Video BW
1 kHz
Figure
,
Preset (Factory)
,
. Set the analyzer by pressing the following
Off
,
10 dB
,
3 kHz
Functional Testing
Scale Fidelity
Recommended
HP/Agilent Model
11852B
1250-1534
3-5.
,
on the
Preset
System
,
3-41