Figure 3-7

Procedure

1. Calibrate the tracking generator by pressing
2. Press
3. Zero and calibrate the power meter and RF power sensor. Make sure
75 Ω RF Out only: Zero and calibrate the 75 Ω power sensor.
NOTE
4. Connect the 100 kHz to 1.5 GHz power sensor to the RF Out on the
75 Ω RF Out only: Connect the 75 Ω power sensor through an adapter
NOTE
to the RF Out 75 Ω.
5. Press REL on the power meter. The power meter readout
6. Set the analyzer center frequency to 100 kHz.
75 Ω RF Out only: Set the analyzer center frequency to 1 MHz.
NOTE
Chapter 3
Tracking Generator Level Flatness: HP E4401B and E4411B, Options 1DN
,
. Connect the RF Out to the Input when prompted.
Align Now
TG
,
System
Power On/Preset
analyzer, then wait for the preset routine to finish. Set the analyzer
by pressing the following keys:
,
FREQUENCY
50 MHz
,
CF Step Auto Man
500 MHz
,
SPAN
Zero Span
Source, Amplitude (On)
,
Source
Amplitude (On)
the power meter is reading out in dBm. Enter the power sensor
5 MHz cal factor into the power meter.
analyzer as shown in
amplitudes are now relative to the power level at 50 MHz.
,
Preset (Factory)
(50 Ω RF Output only)
,
0 dBm
(75 Ω RF Output only)
,
+42.76 dBmV
Figure
3-7.
Functional Testing
and 1DQ
,
,
System
Alignments
,
on the
Preset
3-47