Figure 3-3
Procedure
1. Zero and calibrate the power meter and power sensor as described in
2. Connect the equipment as shown in
Use only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
CAUTION
connectors, or damage to the connectors will occur.
3. Set the synthesized sweeper controls as follows:
4. Press
Chapter 3
the power meter operation manual.
input.
,
FREQUENCY
50 MHz
,
AMPLITUDE
–8 dBm
,
System
Power On/Preset
analyzer and wait for the preset routine to finish. Set the analyzer
by pressing the following keys.
,
FREQUENCY
50 MHz
,
CF Step Auto Man
50 MHz
,
SPAN
20 kHz
,
AMPLITUDE
–10 dBm
,
,
AMPLITUDE
More
Amptd Units
,
AMPLITUDE
–10 dBm
,
Scale/Div
2 dB
,
BW/Avg
10 kHz
Figure 3-3
,
Preset (Factory)
(75 Ω RF Input Only)
,
dBm
,
Attenuation Auto Man
Functional Testing
Frequency Response
for a 50 Ω analyzer
,
on the
Preset
,
10 dB
3-29