Ecotron - Incubation Shaker
Spare Parts
10.5 Replacing the Liner on the Table
Procedure
Page 76 of 90
The liners on the bottom of the table hold and guide the table. Dirty
or damaged liners can result in the equipment not running
smoothly and cause vibrations. Dirty or damaged liners are easy to
replace. To do so, proceed as follows:
1.
Turn off equipment and pull out the plug.
2.
Open the door to the incubation chamber.
3.
If a tray is installed, remove it by loosening the four Phillips-
head screws (M4x6).
4.
Remove the table by loosening the four hexagonal socket
screws (M6x12).
5.
Turn the table 180° and remove the defective liners by loosen-
ing the Phillips-head screws (M4x10).
6.
Install new liner and reinstall the table.