Advertisement

Error during temperature
measurement.
Heater fan not
working.
Alarm message
Possible cause
The parameter has recently
been changed, and the re-
quired setpoint has not been
reached yet.
The ambient temperature
does not meet the require-
ments.
Error during temperature
measurement.
Heater fan not
working.
Alarm message
Possible cause
The load is too heavy, making
the drive too weak or causing
it to overheat.
The drive belt is torn.
The motor is defective.
03 May 2018
Check the function of the temperature sensor. In
case of a defect, contact your local Infors repre-
sentative.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Meaning
Temperature too high (actual value > setpoint + 1.5°C).
Measures
No action required. The alarm message will dis-
appear as soon as the setpoint is reached.
Check the ambient temperature on the equip-
ment and adjust if necessary.
Check the function of the temperature sensor. In
case of a defect, contact your local Infors repre-
sentative.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Meaning
Rotation speed too low.
Measures
Reduce the load.
If the drive has overheated, the process can be
continued after it has cooled down.
Contact your local Infors representative and have
the drive belt replaced.
Contact your local Infors representative.
Ecotron - Incubation Shaker
Rectifying Faults
Operator
Operator
Personnel
Operator
Operator
Operator
Personnel
Operator
Infors service techni-
cian or licensed dealer
Infors service techni-
cian or licensed dealer
Page 63 of 90