Ecotron - Incubation Shaker
Rectifying Faults
Alarm message
Possible cause
Error when measuring the ro-
tation speed.
The drive control system is
defective.
The motor is defective.
Alarm message

8.2 Fault Tables

Fault
After activating the main switch, the view boxes do not light up.
Possible cause
Power supply to equipment in-
terrupted.
The equipment fuse is defec-
tive.
Fault
The interior lighting does not work.
Possible cause
The lighting element of the in-
terior lighting is defective.
Page 64 of 90
Meaning
The rotation speed is too high.
Measures
Switch the equipment off and then on again.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Contact your local Infors representative.
Meaning
The equipment door is open.
The tables below describe potential faults that do not usually trig-
ger an error message on the display or – with a few exceptions –
an alarm signal. In many cases, operators can resolve these faults
themselves.
General faults
Measures
Check if the plugs are plugged in correctly.
Check the mains connection.
Replace equipment fuse (see chapter 10.1, page
72).
Measures
Replace the lighting element (see chapter 10.3,
page 74).
Personnel
Operator
Operator
Infors service techni-
cian or licensed dealer
Personnel
Operator
Operator
Personnel
Operator