Maintenance and Cleaning

9 Maintenance and Cleaning

9.1 Maintenance

Interval
Maintenance work
Prior to each
Check all seals on the equipment, especially on the door, and re-
use
place if necessary.
Check whether the interior lighting works; have lamps replaced if
necessary.
After each use
Clean equipment; if necessary, disinfect thoroughly.
Annually
Have sensors calibrated at least once per year to ensure that the
measuring results remain accurate.
Every 3 years
If using the "cooling" option, have the coolant replaced. To do this,
contact your local Infors representative.
Page 68 of 90
DANGER
To prevent life-threatening electric shocks, always switch off the
equipment and disconnect it from the power supply before carry-
ing out any maintenance or cleaning.
The equipment requires hardly any maintenance. This reduces the
running costs to certain regular checks and cleaning. The following
table describes the maintenance work that is required to ensure
optimum, fault-free operation.
If increased wear is detected during the regular checks, the re-
quired maintenance intervals must be reduced to suit with the ac-
tual signs of wear. Keep in mind that various media or gases will
have more or less corrosive effects on the metal parts. When using
particularly aggressive substances, more frequent checks are re-
quired in order to maintain smooth equipment operation.
Contact the manufacturer if you have any questions regarding
maintenance work or intervals; see Service contact details on page
2.
To be carried out
by
Qualified techni-
cians
Qualified techni-
cians
Operator
Qualified techni-
cians
Infors service tech-
nician or licensed
dealer
03 May 2018