Maintenance and Cleaning

8 Maintenance and Cleaning

8.1 Maintenance

Interval
Maintenance work
Prior to each use Check all seals on the device, especially on the door, and replace
if necessary.
Check whether the interior lighting works; have lamps replaced if
necessary.
After each use
Clean the device; if necessary, disinfect thoroughly.
Annually
Have sensors calibrated at least once per year to ensure that the
measuring results remain accurate.
Every 3 years
If using the cooling option, have the coolant replaced. To do this,
contact INFORS HT.
Page 92 of 114
WARNING
To prevent life-threatening electric shocks, always switch off the
device and disconnect it from the power supply before carrying
out any maintenance or cleaning.
The device requires hardly any maintenance. This reduces the running
costs to certain regular checks and cleaning. The following table de-
scribes the maintenance work that is required to ensure optimum,
fault-free operation.
If increased wear is detected during the regular checks, the required
maintenance intervals must be reduced to suit with the actual signs
of wear. Keep in mind that various media or gases will have more or
less corrosive effects on the metal parts. When using particularly ag-
gressive substances, more frequent checks are required in order to
maintain smooth device operation.
Contact the manufacturer if you have any questions regarding
maintenance work or intervals; see Service contact details on page 2.
To be carried out by
Qualified technicians
Qualified technicians
Operator
Qualified technicians
INFORS HT service tech-
nician or licensed dealer