Blood Bag Cups; Rotor Labware Assemblies - Beckman Coulter Avanti J Series Instructions For Use Manual

Rotors and tubes
Table of Contents

Advertisement

Tubes, Bottles, and Accessories

Blood Bag Cups

Blood Bag Cups

Rotor Labware Assemblies

Plug
Air-Vent
Cup Cover
Filter
Spout
HarvestLine
System Liner
Green Gasket
(369261)
Cup
2-10
Polypropylene blood bag cups are available for use in swinging-bucket
rotors to obtain cell-free plasma for cell packing or for leukolyte
depletion. Different sizes of cups are available to accommodate single,
double, triple, or quad pack blood bags. Refer to the applicable rotor
manual to determine the correct blood bag cup to use. Blood bag cups are
autoclavable.
Blood bags should be loaded into the cups outside of the centrifuge to
avoid tripping the centrifuge imbalance monitor during loading.
The JS-5.0 labware assembly has an available HarvestLine System liner. If
liners are not used, the sample can be loaded directly into the cup and a
partition can be inserted to minimize sample disturbance at low g forces.
The gasket and the cup and cover surfaces that contact the gasket must be
dry to ensure sealing. Gasket 36926 (green) is used when liners are used;
gasket 369257 (red) is used when the cup is used alone, with or without a
partition. The cup cover top surface can be written on to identify the
assembly or sample.
The HarvestLine System for the JLA-8.1000 and JLA-9.1000 rotors provides
a convenient method of loading, recovering, and storing samples run in
these rotors. Up to six rotor bottles are placed in the filling rack, and a
liner is placed into each bottle. The liners are loaded with sample through
a funnel or fermentor hose. The valve in the neck of each liner is then
sealed and the liner necks folded to fit inside the bottles. The bottles are
sealed with rotor plugs and cap/closures, and the sealed bottles are placed
into the rotor canisters for centrifugation. After centrifugation, the liner
valves are cut off and the supernatant decanted, either for storage or
disposal. The liners can then be heat-sealed for pellet storage or disposal.
PN JR-IM-10AG

Hide quick links:

Advertisement

Table of Contents
loading