Labware - Beckman Coulter Avanti J Series Instructions For Use Manual

Rotors and tubes
Table of Contents

Advertisement

Using Swinging-Bucket Rotors
Description
Table 5.1 General Specifications for Beckman Coulter J Series Swinging-Bucket Rotors (Continued)
Rotor Profile and Name
JS-4.2
JS-4.2SM
(use only in
J6 series
centrifuges)
JS-4.2A
JS-4.2SMA
(use only in
J6 series
centrifuges)
JS-4.0
JS-3.0
JS-2.9
a. Maximum speeds are based on a solution density of 1.2 g/mL.
b. Critical m speeds are based on a solution density of 1.2 g/mL.
† Critical speed range is the range of speeds over which the rotor shifts so as to rotate about its center of mass. Passing through or running at the
critical speed range is characterized by some vibration.

Labware

Swinging-bucket rotors can accommodate a variety of tubes, bottles, multiwell titer plates,
and blood bags, listed in individual rotor manuals. Refer to
sealing requirements. Observe the maximum rotor speeds and fill volumes listed in the applicable
rotor manual.
5-4
Critical
Radial Distances (mm)
a
Speed
Max Speed
/
b
RCF/
Range
r
k Factor
(rpm)
max
4200 rpm
600
254
5020 × g
to
11,502
800
4200 rpm
600
248
4900 × g
to
800
4200 rpm
600
254
5020 × g
to
11,502
800
4200 rpm
600
248
4900 × g
to
800
4000 rpm
600
226
4044 × g
to
15,296
800
3000 rpm
600
254
2560 × g
to
22,548
800
2900 rpm
600
265
2500 × g
to
24,400
800
Number of
Tubes x Nominal
Capacity of
r
r
Largest Tube
av
min
6 × 1 liter
184
114
6 blood bags
18 microplates
336 RIA tubes
182
116
6 triple or
quad pack
blood bags
6 × 1 liter
184
114
6 blood bags
18 microplates
336 RIA tubes
182
116
6 triple or quad pack
blood bags
4 × 1 liter
156
86
4 blood bags
12 microplates
148 RIA tubes
6 × 1 liter
184
114
6 blood bags
18 microplates
336 RIA tubes
12 × 500 mL
192
118
blood bags
CHAPTER 3
for tube filling and
Nominal
Rotor
Capacity
6 liters
6 liters
6 liters
6 liters
4 liters
6 liters
6 liters
PN JR-IM-10AG

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents