Mechanical Safety - Beckman Coulter Avanti J Series Instructions For Use Manual

Rotors and tubes
Table of Contents

Advertisement

Safety Notice

Mechanical Safety

the need for aerosol protection. Handle other infectious samples according to good laboratory
procedures and methods to prevent spread of disease. Because spills may generate aerosols,
observe proper safety precautions for aerosol containment. Do not run toxic, pathogenic, or
radioactive materials in the rotor without taking appropriate safety precautions. Biosafe
containment should be used when Risk Group II materials (as identified in the World Health
Organization Laboratory Biosafety Manual) are handled; materials of a higher group require more
than one level of protection.
• Dispose of all waste solutions according to appropriate environmental health and safety
guidelines.
• If disassembly reveals evidence of leakage, you should assume that some fluid escaped the
container or rotor. Apply appropriate decontamination procedures to the centrifuge, rotor,
and accessories.
Mechanical Safety
• Use only the rotors, components, and accessories designed for use in the rotor and centrifuge
being used (refer to the applicable rotor manual). The safety of rotor components and accessories
made by other manufacturers cannot be ascertained by Beckman Coulter. Use of other
manufacturers' components or accessories in Beckman Coulter rotors may void the rotor warranty
and should be prohibited by your laboratory safety officer.
• Rotors are designed for use at the speeds indicated; however, speed reductions may be required
because of weight considerations of tubes, adapters, and/or the density of the solution being
centrifuged. Be sure to observe the instructions in the applicable rotor manual.
• NEVER attempt to slow or stop a rotor by hand.
• The strength of containers can vary between lots, and will depend on handling and usage. We
highly recommend that you pretest them in the rotor (using buffer or gradient of equivalent
density to the intended sample solution) to determine optimal operating conditions. Scratches
(even microscopic ones) significantly weaken glass and polycarbonate containers.
To help prevent premature failures or hazards by detecting stress corrosion, metal fatigue, wear or
damage to anodized coatings, and to instruct laboratory personnel in the proper care of rotors,
Beckman Coulter offers the Field Rotor Inspection Program (FRIP). This program involves a visit to
your laboratory by a specially trained Beckman Coulter representative, who will inspect all of your
rotors for corrosion or damage. The representative will recommend repair or replacement of at-
risk rotors to prevent potential rotor failures. Contact your local Beckman Coulter office to request
this service.
It is your responsibility to decontaminate the rotors and accessories before requesting service by
Beckman Coulter Field Service.
vi
PN JR-IM-10AG