Microprocessor-Controlled Centrifuges - Beckman Coulter Avanti J Series Instructions For Use Manual

Rotors and tubes
Table of Contents

Advertisement

Microprocessor-Controlled Centrifuges

Operating temperatures for most rotors are contained in memory in microprocessor-controlled
centrifuges (model J6-MI, and discontinued models J6-MC, J2-MI, J2-21M, J2-MC, and J2-21M/E). Set
temperature compensation as follows for rotors not in centrifuge memory.
1
Press the
required sample temperature.
2
Find the compensation value in
corresponds with the required temperature and run speed.
3
Press
COMP ADJ
The word "
4
Use the keypad to enter the compensation value.
Press the ±. key to enter a minus sign; pressing it again will remove the minus sign.
5
Check the temperature display.
If the entry is incorrect, press
6
When the entry is correct, press
PN JR-IM-10AG
key on the centrifuge control panel and then use the keypad to enter the
TEMP
APPENDIX B
.
" flashes in the
COMP
TEMPERATURE
and reenter the digits.
ce
ENTER/RECALL
General Operating Information
(or in the applicable rotor manual) that
display and the display flashes.
.
1
Rotors
1-21