Operation
Settings
5 - 54

5.4.3 Recorder function, Record of measured values

With the Record of measured values function, the recorder output pro-
vides a current that is independent of the measured value. With the
Recorder, Recorder Beginning, Recorder End Point and di/dt parame-
ters, the output of measured values is defined.
Setting
Recorder
Recorder Beginning
Recorder End Point
di/dt
Response with stop
Note
The setting of the recorder parameters is described in section
5.4.2 S
ETTING UP THE RECORDER

5.4.4 Recorder function, Controller

Selection/Values
0 - 20 mA
4 - 20 mA
within the measu-
ring range
(depending on the
analyzer module)
0.1 mA/s
1 mA/s
5 mA/s
10 mA/s
20 mA/s
any in the range
0 ... 20 mA
Hold
.
TresCon UNO
Explanation
Signal range of the
recorder output
Note:
Minimum distance:
10% of the measuring
range
Attenuation: Speed of
change of the output cur-
rent in the case of an
erratic change of the
input signal.
In the Stop mode, the
current output provides
the defined current
value. Also in the case of
AutoCal, AutoClean etc.
The current at the output
remains unchanged.
ba43134e08
05/2017