Installing The Permeate Switch - wtw TresCon UNO A111 Operating Manual

Table of Contents

Advertisement

Installation
Pre-assembling the
component
3 - 30

3.7.1 Installing the permeate switch

The permeate switch is screwed to the mounting element of the
TresCon UNO instead of the series overflow vessel.
Replace the triple overflow vessel (left) of the permeate switch by the
simple overflow vessel of TresCon UNO (Fig. 3-25).
3
P r
o b
e 2
P r
o b
e 1
1
Fig. 3-25 Pre-mounting the permeate switch
1
Pull all tube endings (pos. 1) off the overflow vessel.
2
Unscrew the two knurled-head screws (pos. 2) and remove the
lid of the overflow vessel.
3
Unscrew the overflow vessel from the mounting plate.
4
Screw the simple overflow vessel on the mounting plate and
close it with the lid.
5
Attach labels to the overflow vessels (pos. 3): left = "Test
sample 2", right = "Test sample 1".
2
ba43134e08
TresCon UNO
05/2017

Advertisement

Table of Contents
loading

Table of Contents