Installation

3.1
Scope of delivery
The following articles are included within the scope of delivery:
 TresCon UNO basic instrument, equipped with analyzer module
 Mounting element for wall mounting
 Tray of reagents
 Overflow vessel with tubing material
 HT drain set DN40 with connection piece for overflow vessel
 Chemicals for the operation of the analyzer module
 TresCon UNO operating manual for basic instrument and analyzer
module
3.2
Prerequisites for installation
A waste water connection that complies with the regulations of the local
water supply must be present at the site where the instrument is loca-
ted.
The sample is fed via nontransparent tubes with a recommended inter-
nal diameter of 1.5 mm.
Before connecting the instrument, ensure that the values of the mains
power plug agree with the specifications in chapter 7 T
of this operating manual. The mains power plug must have a protective
earth connection.
The instrument is intended for operation indoors. Avoid direct solar
radiation. When selecting the mounting wall, the weight of the instru-
ment must be taken into consideration (see chapter 7 T
The wall and mounting material must fulfill the requirements specified
in chapter 7 T
ECHNICAL DATA
any bright light source to ensure an optimum image quality of the dis-
play. The space it requires is indicated by the dimensions marked on
the instrument views on the following pages.
The type and quantity of the reagents required for operation depend on
the analyzer module.
Note
The consumption of reagents depends on the intervals selected for
measuring and cleaning. For more detailed information, please refer to
the operating manual of the analyzer module.
. Avoid installing the instrument opposite
Installation
ECHNICAL DATA
).
ECHNICAL DATA
3 - 1