Appendix
PROFIBUS connection
PROFIBUS record (PBP)
8 - 20
8.4.3 Commands for PROFIBUS
PC
Profibus DP
PKV 30-DPS
PBP overhead
1 byte
1 byte
Handshake Slave address
RS485
Handshake
(binary)
Slave address
(binary)
Length information
(binary)
ASCII message
Note
Further information on wiring, configuration and the record can be
found in the "PKV 30-DPS instrument manual" and the manual "Con-
nection of WTW measured value recording systems to the PROFIBUS-
DP" that are enclosed in the PKV 30-DPS converter.
WTW monitor
with RS 485 option
RS 485
PBP message
64 bytes (fixed length)
1 byte
Length information
n ASCII
this byte must be toggled (1<->0),
in order to trigger a data transfer on the
RS485 side
RS485 address of the measuring transmitter
(1...31)
Length of the ASCII message in bytes (1...61)
The return values in the response are separa-
ted by spaces.
The length of the response is variable.
TresCon UNO
TresCon
PBP overhead
n bytes
ASCII
message
ba43134e08
05/2017