TresCon UNO
Setting up output
currents
ba43134e08
05/2017

6.1.1 Testing the recorder

You can select default values for the recorder outputs. The currents
selected are immediately applied to the outputs as soon as they are
selected.
Parameter
I1 (output 1)
I2 (output 2)
I3 (output 3)
1
Measuring mode
Call up the Service menu: Simultaneously press the keys,
4 and 2. The request Enter PIN appears.
Input the PIN digit-by-digit with 68. Change the digit
2
with 57 and confirm with 3. The Service Mode
menu appears.
Select the Recorder menu item with 68 and confirm
3
with 3. The Recorder service mode appears.
Select the parameters with 68 and switch into input
4
mode with 3.
Enter the required test current with 68. Change to the
5
next digit with 57 and confirm with 3.
Values
0 to 20 mA
0 to 20 mA
0 to 20 mA
Maintenance
6 - 3