Output Format Of Measured Values - wtw PhotoLab Spektral Operating Instructions Manual

Table of Contents

Advertisement

16 RS 232 C interface
Command
CSPK [Lmin Lmax]
CEME
REME [MMM]
CCLR [MMM]
The error message, Invalid command ap-
pears if commands are unknown or cannot
be carried out (e. g. if optional parameters do
not agree with the cell coding). Optional pa-
rameters [MMM] and [LLL] need only be in-
put for uncoded cells.
16.3

Output format of measured values

Character
3
5
6
17
4
9
10
12
4
4
Data fields are separated by spaces.
Character set: IBM, code page 437
Meaning of the special characters:
Character
!
t/T
*
Q
108
Function
Measurement and transmission of all measured values of the spectrum
[Lmin Lmax] = wavelength range
Input user-defined methods (see chapter
Output user-defined methods
[MMM] = method number
Erase user-defined methods
[MMM] = method number
Meaning
consecutive number (not required for interface commands CMES, CEXT and
CBLA)
method designation
ID number
date and time
special character
measured value
unit
citation form
AQA I. D. (AQA1/AQA2)
dilution factor
Meaning
Measuring with blank value (concentration) or reference absorbance (absor-
bance)
Measuring with turbidity correction / with high turbidity
Measured value outside the measuring range
AQA measurement
13 U
-
SER
DEFINED METHODS
)