The FFT menu table
Menu
Enable
Operate
Source
Window
1/2
2/2
Scale
Display
2/2
Note:
1. Signals that have a DC component or offset can cause incorrect FFT waveform
component magnitude values. To minimize the DC component, choose AC Coupling
on the source signal.
2. To reduce random noise and aliases components in repetitive or single-shot events,
set the oscilloscope acquisition mode to average.
3. To display FFT waveforms with a large dynamic range, use the "dBVrms" scale. The
"dBVrms" scale displays component magnitudes using a log scale.
FFT Window
The
ll
provides four FFT windows. Each window is a trade-off between
frequency resolution and amplitude accuracy. It depends on the desired
Figure 2-14 The FFT menu
Setting
ON
Turn on FFT
OFF
Turn off FFT
FFT
Fast Fourier Transform
CH1
Define CH1 or CH2 as FFT source
CH2
Rectangle
Hanning
Select window for FFT
Hamming
Blackman
Go to the next menu page
Back to the previous menu page
Vrms
Set "Vrms" as vertical unit
dBVrms
Set "dBVrms" as vertical unit
Full
Display waveform in full screen
Split
Display waveform in split screen
Back to the previous menu page
Description