:NUMeric:HOLD
Function
Sets or queries the on/off (hold/release) status of
the numeric data hold feature.
Syntax
:NUMeric:HOLD {<Boolean>}
:NUMeric:HOLD?
Example
:NUMERIC:HOLD ON
:NUMERIC:HOLD? -> :NUMERIC:HOLD 1
Description • If :NUMeric:HOLD is set to ON before
:NUMeric[:NORMal]:VALue? or
:NUMeric:LIST:VALue? is executed, all the
numeric data at that point in time can be held
internally.
• As long as :NUMeric:HOLD is set to ON,
numeric data is held even when the numeric
data on the screen is updated.
• For example, if you wish to retrieve various
types of numeric data from each element
at the same point in time, use the following
commands:
:NUMeric:HOLD ON
:NUMeric[:NORMal]:ITEM1 U,1;ITEM2 I,1;...
(Set the numeric data items of element 1.)
:NUMeric[:NORMal]:VALue?
(Receive the numeric data of element 1.)
:NUMeric[:NORMal]:ITEM1 U,2;ITEM2 I,2;...
(Set the numeric data items of element 2.)
:NUMeric[:NORMal]:VALue?
(Receive the numeric data of element 2.)
:NUMeric[:NORMal]:ITEM1 U,3;ITEM2 I,3;...
(Set the numeric data items of element 3.)
:NUMeric[:NORMal]:VALue?
(Receive the numeric data of element 3.)
:NUMeric:HOLD OFF
• If :NUMeric:HOLD is set to ON after having
already been set to ON before, the numeric
data is cleared, and the most recent numeric
data is held internally. When retrieving numeric
data continuously, this method can be used to
circumvent the need to repeatedly set
:NUMeric:HOLD to OFF.
IM WT310E-17EN